i7 Reflection Port During Transmission Measurement

Q - Should there be anything in the reflection view port when taking transmission measurements?

A - Yes.  The Color i7 transmission kit includes a transmission port plate and a white spectralon disk to be used for transmission calibrations and measurements.  The transmission port plate is to be used when doing transmission measurements other than Haze or Reflectance/Transmission.  When measuring Haze or Reflectance/Transmission the LAV port plate is used with the transmission spectralon disk at the reflectance port.